Feature:
1. Double-beam optical system.
2. The Main unit and PC software can
independently implement functions of Quantitative; Kinetics; Wavelength Scan;
Multi Wavelength; DNA/Protein and
Data Printing, PC software can complete the function of data processing.
3. Suspended posture optical system design,
strengthen and thicken the bottom plate to eliminate the
vibration or transformation’s
impact on the optical system.
4. 24-bit high speed and high precision A/D
conversion, and improve the sensitivity of the instrument.
5. 0.5/1.0/2.0/4.0/5.0 bandwidth can be
adjusted automatically
6. The core components are imported with original packaging.
7. The best optical system, based on top
structure design, top technological requirements and top raw materials.
8. Standard with PC software.
Characteristics:
1. Double-beam optical system.
2. The Main unit and PC software can independently implement
functions of Quantitative; Kinetics; Wavelength Scan;
Multi Wavelength;
DNA/Protein and Data Printing, PC software can complete the function of data
processing.
3. Suspended posture optical system design, strengthen and
thicken the bottom plate to eliminate the
vibration or transformation’s impact
on the optical system.
4. 24-bit high speed and high precision A/D conversion, and
improve the sensitivity of the instrument.
5. 0.5/1.0/2.0/4.0/5.0 bandwidth can be adjusted automatically
6. The core components are imported with original packaging.
7. The best optical system, based on top structure design, top
technological requirements and top raw materials.
8. Standard with PC software.
Technical data:
Wavelength Range
|
190-1100nm
|
Bandwidth
|
0.5/1/2/4/5nm
|
Wavelength Accuracy
|
±0.1nm(D2 656.1NM),±0.3nm
|
Wavelength Reproducibility
|
≤0.1nm
|
Photometric Accuracy
|
±0.2%T
|
Photometric Repeatability
|
≤0.15%T
|
Straylight
|
≤0.03%T
|
Stability
|
±0.0004A/h(500nm)
|
Baseline Flatness
|
±0.0001A
|
Noise
|
±0.0005A
|
Photometric Range
|
0-200%T, -4.0--4.0A, 0-9999C
|
Wavelength setting mode
|
Automatic
|
Scanning speed
|
HI,MED,LOW,Optional
|
Output
|
USB Port
|
Printer port
|
Parallel Port
|
Display
|
LCD(320*240)
|
Light Source
|
Deuterium&Tungsten Halogen
Lamp
|
Detector
|
Silicon Photodiode
|
Power
|
AC 220V/50Hz or 110V/64Hz
|
Dimension
|
625*430*210mm
|
Weight
|
28Kg
|