Large Steam Sterilizer Class B Sterilization

Large Steam Sterilizer Class B Sterilization


Feature
Sterilization temperature 121°C to 135°C, EN-285 class B Sterilization cycle
Pre vacuum-water filling-heating-sterilizing-ventilation-post vacuum drying-end cycle

Pre vacuum air removal and Post vacuum Drying system
Pre-vacuum for effective air removal that results in increasing sterilization performance.
Post-vacuum for drying with remaining steam removal

Full automatic operation system with Microprocessor PID program controller
Easy operation with full automatic process cycle operated by microprocessor controller from pre-programmed process cycle

Self-diagnostic and Safety system
In case of any malfunction, it alarms and displays error codes for easy maintenance, being stop working with automatic ventilation process for safety purpose

Air filtration cooling system
For the process of drying cycle, 0.2µm HEPA filtered clean air flows into the chamber in order to protect air contamination and to cool down below 60°C

Built-in steam generator and water reservoir
Water re-circulation system reduces the total sterilizing water consumption and pre heated double jacket chamber stand – by for quick operation

Automatic door locking system
Door can be easily sliding up and down to open or close by counter balanced weighing mechanism. Simply  press start key to operate, door will be automatically locked and unlocked by mechanical can lock system. During operation cycle, door cannot be opened

Safety Mechanism
-        Over pressure release valve
-        Over temperature protector with Thermostat
-        Higher temp set protector by Digital high alarm limit
-        Visible and Audible alarm system

-        Automatic dual door locking system
Mode
Temperature
(°C)
Pressure
(kgf/cm²)
Sterilizing Time (min)
Drying Time
(min)
Unwrapped 1
121
1.1
20
15-20
Unwrapped 2
134
2.1
5
15-20
Wrapped 1
121
1.1
30
20-25
Wrapped 2
134
2.1
15
20-25
Liquid
121
1.1
20
0
Bowie Dick
134
2.1
4
5-10
User Program
110-135
0.4-2.3
4-30
0-60

MODEL
LAC-7305SP
LAC7505SP
Dimensions                      Outer
(W x D x H mm)          Inner
700x1140x1745
540x1048x540
800x1240x1845
640x1280x640
Chamber                      Capacity
                                       Type
                                       Material
300ℓ
500ℓ
Double Jacketed Square chamber
Stainless steel 304 (STS 316 L option)
Steam Generator        Built-in
45ℓ Boiler / 12 kW Heater
60ℓ Boiler / 18kW Heater
Stainless Steel Tray
480x980 / 340x980 mm
580x1180 / 440x1180 mm
Sterilization               Temp. Range
                                Pressure Range
                                               Course
121°C - 135°C (Max)
1.2 – 2.3 Kgf/cm²
Liquid /Unwrapped 1 & 2 /Wrapped 1 & 2 /B/D test /User Program
Vacuum
Pre-vacuum with water ejector vacuum pump system
Controller
6 Programmed with 1 User  Program in Digital PID Controller
Display
LCD monitor for Temperature, Pressure, Timer, Program cycle
Door
Single Door /Auto vertical sliding door, Automatic Door Locking System
Printer
Built-in Thermal Printer
Electric Supply
380V, 50/60Hz 3Ph, 4 Line














MODEL
LAC-7307DSP
LAC-7505DSP
Door / steam system
Pass – Through Front & Rear Double Doors / Built – in steam generator
MODEL
LAC-7306ES
LAC-7506ES
Steam System Supply
External steam supplied models without steam generator